[abstract] REVIEW OF IMPLANTED CARDIAC RHYTHM DEVICES IN THE HYPERBARIC CHAMBER

Rubicon Research Repository/Manakin Repository

[abstract] REVIEW OF IMPLANTED CARDIAC RHYTHM DEVICES IN THE HYPERBARIC CHAMBER

Show full item record


Title: [abstract] REVIEW OF IMPLANTED CARDIAC RHYTHM DEVICES IN THE HYPERBARIC CHAMBER
Author: Simmons, S; Zmaeff, J
Abstract: Background: Implanted pacemakers and AICDS are getting smaller. There have been recent unpublished anecdotal reports of pacemaker failure related to pressurization. There is no centralized source of current information for anticipating the effect of pressure on these devices. Method: Manufacturers of cardiac pacemakers and implanted defibrillators were contacted for the results of testing and updated recommendations for pressure limits on their devices. Chart review was done on patients with pacemakers or AICDS who underwent hyperbaric treatments. Results: An updated table showing maximum tested depths and information about different cardiac devices. Conclusions: No adverse events noted at this chamber over the surveillance period. Most implanted devices are rated to at least 2.4 ATA.
Description: Undersea and Hyperbaric Medical Society, Inc. (http://www.uhms.org )
URI: http://archive.rubicon-foundation.org/1072
Date: 2001

Files in this item

Files Size Format View
abstract.txt 259bytes Text file View/Open

This item appears in the following Collection(s)

  • UHMS Meeting Abstracts
    This is a collection of the published abstracts from the Undersea and Hyperbaric Medical Society (UHMS) annual meetings.

Show full item record